HP 8924C Reference Manual page 274

Cdma mobile station test set
Hide thumbs Also See for 8924C:
Table of Contents

Advertisement

Chapter 4, Description of Fields
Fields That Begin with the Letter I
IF Filter
If Unit-Under-Test Fails
Init Power
274
Screens Where Field is Present
CDMA CALL CONTROL
CDMA TRANSMITTER POWER RANGE TEST
This field selects the desired IF filter bandwidth for modulated signals being
analyzed.
Screens Where Field is Present
TX TEST
DUPLEX TEST
RF ANALYZER
This field selects what happens if the measured results do not meet criteria for
passing the current test. The underlined entry is the active choice.
Continue continues with testing even though a test in the test sequence fails to meet
its test specification limits. When this occurs, an error is listed on the test-results print-
out, is displayed on the CRT, or both.
Stop stops the test when a failure occurs and requires operator intervention before test-
ing proceeds.
Screens Where Field is Present
TESTS (Execution Conditions)
This field allows entry of the initial power offset for system access (INIT_PWR).
HP-IB Example
"DISP CCON;CDMA:CELL:CONFigure:INITial:POWer 0"
sets INIT_PWR to 0.
Operating Considerations
This field specifies a correction factor the mobile station adds to the estimated
open-loop output power for setting the power level of the first (initial) access
probe in an access probe sequence.
Screens Where Field is Present
CDMA CELL SITE CONFIGURATION
S:\hp8924c\REFGUIDE\MANUAL\fields.chp

Advertisement

Table of Contents
loading

Table of Contents