HP 8924C Reference Manual page 166

Cdma mobile station test set
Hide thumbs Also See for 8924C:
Table of Contents

Advertisement

Chapter 4, Description of Fields
Fields That Begin with the Letter A
Amplitude
166
This field sets the output power of the Test Sets's transmitter (that is, the output
power of the Test Set's RF Generator). See
Command Syntax chapter of the HP 8924C Condensed Programming Reference
Guide for programming command syntax.
The :AMPLitude command is used to control this field.
To query the current setting of the amplitude field use the :AMPLitude?
command.
Syntax
:AMPLitude <real number> <units>
:AMPlitude?
Example
OUTPUT 714;"CALLP:AMPL -50 DBM"
OUTPUT 714;"CALLP:AMPL?"
ENTER 714;Ampl_val$
Operating Considerations
This is a priority control field. Accessing the RX TEST or TX TEST screen
overrides any changes made to this field in other screens.
If a microphone is connected, and the amplitude is set to
microphone causes the amplitude to turn on to its previous level until the
microphone is no longer keyed.
See Also
"Atten Hold" on page 168
"MIC/ACC" on page 76
Screens Where Field is Present
DUPLEX TEST
RF ANALYZER
RF GENERATOR
RX TEST
SPECTRUM ANALYZER
CALL CONTROL
ANALOG MEAS
"RF Generator "
in the HP-IB
, keying the
Off
S:\hp8924c\REFGUIDE\MANUAL\fields.chp