Max Req Seq, Max Rsp Seq
Max Slot Cycle Index
NOTE:
When i = 7, the value is quite large at 163.84 seconds. You may not want to wait this long
during normal testing.
This field allows simultaneous entry of the maximum number of access probe
sequences for an access channel request and the maximum number of access
probe sequences for an access channel response (MAX_REQ_SEQ or
MAX_RSP_SEQ). These are separate parameters in the CDMA specification but
the Test Set sets both parameters to a single value.
HP-IB Example
"DISP CCON;CDMA:CELL:CONFigure:MAXReqrspseq 3"
sets both parameters to 3.
Operating Considerations
This field indicates the maximum number of access probe sequences sent by the
mobile when the mobile is requesting service (Max Req Seq) or when the mobile
has been paged (Max Rsp Seq). If the Call Limit field on the CDMA Call Site
Configuration screen is set to "Page", the mobile station will transmit the number
of access probe sequences entered in this field.
Screens Where Field is Present
CDMA CELL SITE CONFIGURATION
Max Slot Cycle Index allows you to set the slot cycle used by the HP 8924 to page
a mobile station. It is set by integer values, which correspond to time using the
T
=
following formula:
7 (0 is default).
HP-IB Example
"CDMA:CELL:CONFigure:MSCindex 1"
sets the cycle to 2.56 seconds.
Screens Where Field is Present
CDMA CELL CONFIGURE
S:\hp8924c\REFGUIDE\MANUAL\fields.chp
Fields That Begin with the Letter M
i
×
1.28 2
seconds, where i has integer values from 0 to
Chapter 4, Description of Fields
291