Chapter 1, Description of Keys
DATA FUNCTIONS Keys
Guidelines for Using Numeric Entry Field Data Functions
60
Increment values are set, scaled, and changed for individual numeric entry fields. The
HP-IB data function commands must immediately follow the HP-IB command for the
individual field. For example, to set the increment value for the RF Generator frequency
to 2.5 MHZ, the following command string would be sent to the Test Set:
OUTPUT 714;"RFG:FREQ:INCR 2.5 MHZ"
HP-IB Units (UNITs) are used with the data functions to specify the units-of-measure
for numeric data which is read or set through HP-IB.
Refer to
"To Specify Units-of-Measure for HP-IB Data Transfer"
chapter of
Overview
the HP 8924C User's Guide.
Data function settings are not retained. The setting values are initialized or changed
under the following conditions:
The Test Set is turned off (values initialized on power up).
The Test Set is PRESET (values initialized).
A saved register is recalled (values changed to those in the recalled register).