Chapter 4, Description of Fields
Fields That Begin with the Letter S
Step#
Stop Length
Strength
412
The
field lists the order of selected tests for the current procedure.
Step#
Operating Considerations
Use the knob, up-arrow, or down-arrow key to select the step where you want to
delete, replace, or insert a test.
Screens Where Field is Present
TESTS (Order of Tests)
This field specifies the number of stop bits used for serial communications when
using the rear-panel serial port. Selecting this field displays a list of stop bit
choices.
This setting is maintained after the instrument is turned off.
Screens Where Field is Present
I/O CONFIGURE
This column in the CDMA Mobile Reporting table lists the power level most
recently measured and reported by the mobile station for each pilot the mobile
station is currently tracking.
All numbers displayed in the CDMA Mobile Reporting table reflect the last pilot
strength message received from a mobile station. Before any numbers are
displayed, a call must be connected and Sector B must be transmitting at a high
enough power level to cross either of two thresholds set up by the T_Add and
T_Comp fields.
The CDMA Mobile Reporting table can be cleared by selecting the MS Report
(Clear) field.
HP-IB Example
"DISP CMOB;CDMA:MOB:PIL:STR?"
queries the results displayed in the CDMA Mobile Reporting table.
The value returned by the Test Set includes Status, PN Offset, Strength, and Keep
for each pilot currently being tracked by the mobile station. These values are sent
as a comma-separated list of IEEE floating point numbers, 16 characters each,
ending with a line feed character.