Chapter 1, Description of Keys
Keys That Begin with the Letter I
NOTE:
When querying a field setting through HP-IB, the Test Set always returns numeric values in
HP-IB units or attribute units, regardless of the field's current display units setting. Numeric
values are expressed in scientific notation.
Refer to
chapter of the HP 8924C User's Guide.
NOTE:
Integer-only fields (such as Intensity and Print Adrs) have a fixed increment of 1, and
cannot be changed.
32
"To Specify Units-of-Measure for HP-IB Data Transfer"
To Query the Increment Value Display Units.
commands to query the units-of-measure used to display the increment value.
Refer to
"
To Specify Units-of-Measure for CRT Display"
chapter of the HP 8924C User's Guide for description of display units.
Syntax
:INCRement:DUNits?
Example
OUTPUT 714;"RFG:FREQ:INCR:DUN?"
ENTER 714;Disp_unit$
This queries the increment value's display units for the RF Generator's
frequency.
Programming INCR÷10
Syntax
:INCRement:DIVide
Example
OUTPUT 714;"RFG:FREQ:INCR:DIV"
If the RF Generator's frequency increment is 10 MHz, this command reduces the
increment value from 10 MHz to 1 MHz.
in the Operating Overview
Use the :INCRement:DUNits?
the Operating Overview
in