8.5 Data input commands

8.5.1 Command input format

is made up of 4 characters, CR=0DH, LF=0AH
1
2
3
C1
C2
CR

8.5.2 External taring command

C1
C2
T
SP

8.5.3 External control commands

C1
C2
Code
O
0
4FH
O
1
4FH
O
2
4FH
O
3
4FH
O
4
4FH
O
5
4FH
O
6
4FH
O
7
4FH
O
8
4FH
O
9
4FH
O
A
4FH
O
B
4FH
* While using these commands for remote control purposes, do not press the PRINT button (data
transmission error). In case data transmission error occurs, briefly cut off the weighing scale from its
power supply source.
54
4
LF
Code
54H
30H
No data output
31H
Continuous data output
32H
Continuous output of stable weight value data
Output of stable and unstable weight values after pressing the
33H
PRINT button
Data output upon stable weight value, having previously unloaded
34H
the weighing scale
Data output upon stable weight value. No data output at unstable
35H
weight value. Data output again, after stabilization has been
reached
Data output upon stable weight value. Continuous data outputting
36H
at unstable weight value.
37H
Output of stable weight values after pressing the PRINT button
38H
One-time immediate data output*
39H
One-time data output after stabilization has been reached*
One-time, immediate data output after lapse of a defined period of
41H
time*
One-time, immediate data output after lapse of a defined period of
42H
time and upon stable weight value*
Content
20H
Tare out command
Meaning
EW/EG-N/EWB-e-1627