Agilent Technologies NovoCyte Quanteon Operator's Manual page 129

Flow cytometer
Table of Contents

Advertisement

System Maintenance
The Purge function is only available when the Purge Instrument privilege is
NOTE
enabled. Refer to
privilege.
1
Click Instrument > Fluidics Maintenance > Purge from the NovoExpress
software.
Figure 108.
2
NovoExpress software will prompt a dialog box to guide you through the
purging procedure. Follow the instructions in each step and click OK to move
to the next step until completion.
Figure 109.
3
Power off the instrument. Unplug the power supply and accessories.
4
Sterilize the reagent containers following the procedure described in
Instrument Reagent Containers
5
Pack the instrument and accessories properly for shipment.
If there are operating questions when purging the fluidics system, in order to
avoid damaging the instrument, please stop purging and contact your local
CAUTION
Agilent representative or distributor.
NovoCyte Quanteon Flow Cytometer Operator's Guide
NovoExpress Software Guide
Purge Function in NovoExpress
Software Message When Starting the Purge Procedure
of this guide.
for instructions to enable this
Sterilize
129