System Maintenance
Figure 105.
2
NovoExpress software will prompt a dialog box to guide you through the
decontaminating procedure. Follow the instructions in each step and click OK
to move to the next step until completion.
Figure 106.
Contact your local Agilent representative or distributor for any questions regarding
NOTE
this procedure.

Prime Fluidic System before Long-Term Storage

If there is no plan to run the instrument for more than two weeks, a thorough
cleaning process is recommended to prevent salt crystallization in the fluidic
system.
1
When the instrument is in Ready status, replace all the instrument reagents
with Milli-Q or 0.2 µ m filtered DI water.
NovoCyte Quanteon Flow Cytometer Operator's Guide
Decontamination Function in NovoExpress
NovoExpress Dialog Box to Guide the Decontamination Procedure
127