NovoCyte Quanteon Technical Specifications
Table 15 NovoCyte Quanteon Technical Specifications
Data
Acquisition
144
Sample Flow Rate
Average Sheath Flow Rate
Sample Aspiration Volume
Sample Recovery
Compatibility to Autosampler No fluidic tubing disassembly or re-connection
Fluid Level Sensing
Fluid Container Capacity
Carryover
Sample Injection Probe (SIP)
Rinse
Fluidics System Monitoring
Fluidics Maintenance
Software
Parameters
Dynamic Range
Fluorescence Photodetector
Gain Control
Compensation
File Export
Data Report
Usage Monitor
NovoCyte Quanteon Flow Cytometer Operator's Guide
1
5-120 µ L/min, continuously adjustable
6.5 mL/min
5 µ L -5 mL
Available
required.
Active sensing using weight sensors with
automated warnings when any fluid level is out
of specified range.
3 L sheath, 3 L waste, 500 mL cleaning, 500 mL
rinse. Optional large container for sheath (15 L)
and waste (15 L) with Fluidics Cart.
< 0.1%
Automated flying collar wash of inner and outer
SIP surface after each sampling.
In-line pressure sensor monitors the pressure in
real time. Automated system recovery when
pressure is out of range due to clogging.
Automated startup and shutdown with fluidic
system cleaning. Automated user executable
cleaning, debubble, rinse, unclog, priming and
decontamination. Automated execution of user
definable maintenance sequence.
NovoExpress
Height and Area for FSC, SSC and all
Fluorescent Channels, Width off FSC and Time.
24 bits; 7.2 decades logarithmic scale
0-1000 User adjustable, optimized default gain
setting for each individual channel.
Full inter-beam matrix, during or post
acquisition.
FCS 3.0, FCS 3.1; CSV; Batch PDF report
Automatic report. Customizable report. Batch
PDF report.
Comprehensive Transaction Log and System
Log.