Calibrate NovoCyte Fluidics Station II

NovoCyte Fluidics Station II is calibrated at the factory and should be ready to use
when connected to the instrument after installation. However, in some cases when
there is a need to recalibrate the fluidics station after shipment and movement
(e.g. when a false warning about the liquid level in the reagent containers is
observed), follow the procedures below to recalibrate the fluidics station.
Calibrate Fluidics Station function is only available when Calibrate Fluidics Station
NOTE
privilege is enabled. Refer to
this privilege.
1
Click Instrument > Operation > Calibrate Fluidics Station in the NovoExpress
software.
Figure 110.
2
Follow the instruction in the software, remove the NovoFlow, NovoRinse,
NovoClean, and waste containers from the fluidics station. Click the OK
button.
Figure 111.
3
The software will prompt a dialog box indicating the fluidics station calibration
is successful. Follow the instructions to place the NovoFlow, NovoRinse,
NovoClean, and waste containers back into the fluidics station. Click the OK
button.
130
NovoExpress Software Guide
Calibrate Fluidics Station Function in NovoExpress
Software Message Instructing to Start Fluidics Station Calibration
NovoCyte Quanteon Flow Cytometer Operator's Guide
for instructions to enable