Running Samples
Figure 62. Compensate FITC Spillover into PE Channel in Spillover Matrix
Figure 63. Compensate PE Spillover into FITC, PerCP and APC Channels
Select Preview in the compensation window to visualize the compensation results
NOTE
dynamically when adjusting the spillover matrix coefficient. Click the Expand
Statistics icon
in the bottom-right corner of the plot to show the statistical
calculation results. Look at the corresponding Mean or Median value to judge if the
correct compensation has been achieved.