System Maintenance
Figure 102.
Use lint free wipes to gently wipe of the sheath fluid from the tubing and
CAUTION
avoid sheath fluid dripping onto the instrument.
4
Connect the outlet of the new filter to the female Luer connector, and the inlet
of the new filter to the male Luer connector. Secure the filter onto the filter
rack by gently sliding it through the open slit on the rack.
CAUTION
Ensure the filter is vertically mounted and all the tubing is not twisted.
Under some situation when the Ultrafiltration Filter needs to be removed after initial
NOTE
installation, disconnect the filter from the Luer connectors and connect the male
Luer connector directly to the female one.
NovoCyte Quanteon Flow Cytometer Operator's Guide
Location of the 0.05 µ m Sheath Fluid Ultrafiltration (UF) Filter rack