System Maintenance
NovoFlow sheath fluid inlet filter (1pcs)
NovoRinse solution inlet filter (1pcs)
NovoClean solution inlet filter (1pcs)
Follow the procedure described below to replace all the fluidic system
consumables.
Once all the fluidic system consumables have been replaced with the new ones,
click the Next button to continue.
If Cancel is selected, a dialog box below will be prompted. The process can be
aborted by clicking Yes button.
Figure 92. Software Message for Cancelling Fluidics System Consumable Replacement
A confirmation message as shown below will be prompted. Click Yes button to
inform the system that the consumables have been replaced.
Figure 93. Software Message to Confirm the Fluidics Consumables are Replaced
After confirmation of fluidics consumables being replaced, the NovoExpress
software will prompt a window to remind users to conduct the Priming procedure.
NovoCyte Quanteon Flow Cytometer Operator's Guide