System Maintenance
Figure 85. Software Message Instructing to Load a Sample Tube
Clean the SIP Cleaning Apparatus
1
Turn off the power of the instrument.
2
Prepare a cotton swab soaked with NovoClean solution (1X). Wipe the SIP
cleaning apparatus' lower surface back and forth until it is clean.
Figure 86. Wipe the SIP Cleaning Apparatus' Lower Surface
3
Prepare a damp cloth soaked with Milli-Q or 0.2 µ m filtered DI water. Wipe the
SIP cleaning apparatus back and forth at least five times.
NovoCyte Quanteon Flow Cytometer Operator's Guide