Instrument Operation
9
Ensure all the test information is correct and the sample tube is properly
installed on the tube holder, click Run to start. The software will automatically
adjust the photodetector gain by running the QC particles. When the
adjustment is completed, the software will automatically perform the QC Test.
Click Report when the QC test is completed.
Figure 36. Adjust the Photodetector Gain
Figure 37. Running QC Test
10 Ensure the result shows Pass for all the channels. Click Finish to close the
window and complete the optical configuration modification.
66
NovoCyte Quanteon Flow Cytometer Operator's Guide