Instrument Operation
current user account by following the instructions described in
Privilege
2
Click Instrument > Configuration to open the Instrument Configuration
window. Click Create Copy in the Instrument Configuration window to
generate an editable copy of the configuration. Click Rename to rename the
configuration if needed. Click Change Optical Configuration to continue.
A customized optical configuration can be deleted by clicking Delete. Only the user-
NOTE
defined optical configurations can be deleted.
Figure 28. Customize the Instrument Configuration
3
Open the top cover of the instrument. Insert one end of the Allen wrench
through the hole. Gently hold the proper optical filter or dichroic mirror and pull
it upward to remove it from the slot. Insert the new optical filter or dichroic
mirror back into the slot as shown in the following figure. Close the top cover
of the instrument.