MMS Module
Removal and
Reinstallation
CAUTION
To Remove an MMS Module
1. Set the mainframe line switch to OFF.
2. Remove all rear panel cables going to the module
3. On the bottom of the mainframe front panel is a dark-colored, horizontal
access panel. Pry outward at the top of this panel to open it.
4. With an 8 mm hex-ball driver, loosen the module hex nut latch.
5. Go to the back of the system and press against the module's rear panel
and slide the module forward several inches.
6. From the front of the system, pull the module out.
To Reinstall an MMS Module
1. Set the MMS mainframe line switch to OFF.
2. Check the GPIB address switch on the module for the correct address
setting. Refer to the manual for the MMS module for information on the
HP-MSIB switch. For proper address settings, refer to the system
information chapter.
Reinstalling an MMS module without setting the GPIB address will cause
the system to malfunction. (Not all MMS modules use GPIB settings.)
3. On the bottom of the mainframe front panel is a dark-colored, horizontal
access panel. If necessary, pry outward at the top of this panel to open it.
4. Slide the module into the mainframe.
5. Press against the front of the module while tightening the hex-nut latch
(with an 8 mm hex-ball driver).
6. Close the access panel.
7. Go to the back of the system and connect intermodule cables.
Agilent Technologies E5500 Phase Noise Measurement System A-11
Connector Care and Preventive Maintenance
Removing and Reinstalling Instruments