CAUTION

Required Equipment

NOTE
Absolute Measurement Examples

RF Synthesizer using DCFM

RF Synthesizer using DCFM
This measurement example will help you measure the absolute phase noise
of an RF synthesizer using DCFM.
To prevent damage to the Agilent/HP 70420A test set's hardware
components, the input signal must not be applied to the signal input
connector until the input attenuator has been correctly set for the desired
configuration, as show in
when the Connection Diagram appears.
The following equipment is required for this example in addition the phase
noise test system and your unit-under-test (UUT).
To ensure accurate measurements, you should allow the UUT and
measurement equipment to warm up at least one hour before making the
noise measurement.
Table 7-9
Equipment
Agilent/HP 8663A
Coax Cables
7-48 Agilent Technologies E5500 Phase Noise Measurement System
Table 7-11 on page
7-55. Apply the input signal
Required Equipment for the RF
Synthesizer using DCFM Measurement
Quantity
Comments
1
Must have DCFM Input Port.
Refer to the
"Selecting a Reference"
more information about reference
source requirements
And adequate adapters to connect
the UUT and reference source to
the test set.
Chapter 6,
for