Measurement Difficulties; System Connections - Agilent Technologies E5500A User Manual

Phase noise measurement system
Hide thumbs Also See for E5500A:
Table of Contents

Advertisement

NOTE

System Connections

Residual Measurement Fundamentals

Measurement Difficulties

Measurement Difficulties
Chapter 6, Evaluating Results, contains troubleshooting information to be
used after the measurement has been made, and a plot has been obtained.
When making phase noise measurements it is important to keep your
equipment connected until the measurements have been made, all problems
corrected, and the results have been evaluated to make sure that the
measurement is valid. If the equipment is disconnected before the results
have been fully evaluated, it may be difficult to troubleshoot the
measurement.
The first thing to check if problems occur is the instrument connections and
settings as this is the most common error. It is also important to make sure
the levels are correct into the Agilent/HP 70420A Phase Detector Inputs.
8-28 Agilent Technologies E5500 Phase Noise Measurement System