Method 1, Example 2 - Agilent Technologies E5500A User Manual

Phase noise measurement system
Hide thumbs Also See for E5500A:
Table of Contents

Advertisement

Method 1, Example 2

Figure 12-9
AM Noise Measurement Fundamentals
Method 1: User Entry of Phase Detector Constant
Advantages
Easy method of calibrating the measurement system.
Will measure DUT without modulation capability.
Requires little additional equipment: only a voltmeter or an oscilloscope.
Fastest method of calibration. If the same power levels are always at the
AM detector, as in the case of leveled outputs, the AM detector
sensitivity will always be essentially the same.
Measures the AM detector gain in the actual measurement
configuration. Super-quick method of estimating the equivalent phase
detector constant.
Disadvantages
Has only moderate accuracy compared to the other calibration methods.
Procedure
1. Connect circuit as shown in
the Agilent/HP 70420A Option 001 or Agilent/HP 70427A is available,
use one of the connection diagrams described in
Measurement Block Diagrams" on page
2. Measure the power which will be applied to the AM detector. It must be
between 0 and +23 dBm.
User Entry of Phase Detector Constant AM Noise Measurement Setup
Method 1, Example 2
3. Measure the monitor output voltage on the AM detector with an
oscilloscope or voltmeter. Locate the diode detector's dc voltage along
the bottom of the AM sensitivity graph
Moving up to the diagonal calibration line and over, the equivalent
phase detector constant can then be read from the left side of the graph.
The measured data will be plotted as single-sideband AM noise in
dBc/Hz.
12-10 Agilent Technologies E5500 Phase Noise Measurement System
Figure
12-9, and tighten all connections. If
"AM Noise
12-3.
(Figure 12-8 on page
12-9).

Advertisement

Table of Contents
loading

Table of Contents