Table 7-4
Parameter Data for the Stable RF Oscillator Measurement
Step
Parameters
5
Test Set Tab
• Input Attenuation
• LNA Low Pass Filter
• LNA Gain
• Detector Maximum Input
Levels
Microwave Phase Detector
RF Phase Detector
AM Detector
• Ignore out-of-lock conditions
• Pulsed Carrier
• DC Block
• Analyzer View
• PLL Integrator Attenuation
6
Downconverter Tab
7
Graph Tab
• Title
• Graph Type
• X Scale Minimum
• X Scale Maximum
• Y Scale Minimum
• Y Scale Maximum
• Normalize trace data to a:
• Scale trace data to a new
carrier frequency of:
• Shift trace data DOWN by:
• Trace Smoothing Amount
• Power present at input of DUT
Data
• Auto checked
• Auto checked
• Auto Gain
• 0 dBm
• 0 dBm
• 0 dBm
• Not checked
• Not checked
• Not checked
• Baseband
• 0 dBm
• The downconverter parameters do not apply to this
measurement example.
• Stable RF Oscillator vs Similar Reference Source
• Single-sideband Noise
• 1 Hz
• 10 E + 6 Hz
• 0 dBc/Hz
• - 170 dBc/Hz
• 1 Hz bandwidth
• 1 times the current carrier frequency
• 0 dB
• 0
• 0 dB
Agilent Technologies E5500 Phase Noise Measurement System 7-23