Agilent Technologies E5500A User Manual page 363

Phase noise measurement system
Hide thumbs Also See for E5500A:
Table of Contents

Advertisement

Evaluating Your Measurement Results
Evaluating the Results
For example, applying the 7 dB difference in noise levels, shown in
"Example Comparison of Measurement Results and Reference Source
Noise." on page 15-4
at 10 kHz, to the graph, reveals that the measured
results should be decreased by about 1 dB at 10 kHz to reflect the actual
noise of the UUT.
Figure 15-2
Example Comparison of Measurement Results and Reference Source
Noise.
15-4 Agilent Technologies E5500 Phase Noise Measurement System