Agilent Technologies E5500A User Manual page 56

Phase noise measurement system
Hide thumbs Also See for E5500A:
Table of Contents

Advertisement

CAUTION
Table 5-3
Testing the Agilent/HP 8663A Internal/External 10 MHz
The Agilent/HP 70420A test set's signal input is subject to the following
limits and characteristics:
Agilent/HP 70420A Test Set Signal Input Limits and Characteristics
Frequency
Maximum Signal Input Power
At Attenuator Output, Operating Level
Range:
• RF Phase Detectors
• Microwave Phase Detectors
• Internal AM Detector
• Downconverters:
Agilent/HP 70422A
Agilent/HP 70427A
CAUTION:
To prevent damage to the Agilent/HP 70420A Test Set's hardware components,
the input signal must not be applied to the test set's signal input connector until
the input attenuator (Option 001) has been set by the phase noise software,
which will occur at the connection diagram.
Input Impedance
AM Noise
Agilent Technologies E5500 Phase Noise Measurement System 5-17
Expanding Your Measurement Experience
Limits
50 kHz to 26.5 GHz
+30 dBm
0 to +23 dBm
0 to +5 dBm
0 to +20 dBm
0 to +30 dBm
+5 to +15 dBm
Characteristics:
50 ohm Nominal
dc coupled to 50 ohm load