Agilent Technologies E5500A User Manual page 251

Phase noise measurement system
Hide thumbs Also See for E5500A:
Table of Contents

Advertisement

Figure 8-7
Phase Detector Sensitivity
4. Remove the power meter and reconnect the cable from the splitter to the
Signal Input port.
5. If you are not certain that the power level at the Reference Input port is
within the range shown in the preceding graph, measure the level using
the setup shown in
6. Remove the power meter and reconnect the cable from the splitter to the
Signal Input port.
7. After you complete the measurement set up procedures and begin
running the measurement, the computer will prompt you to adjust for
quadrature. Adjust the phase difference at the phase detector to 90
degrees (quadrature) by either adjusting the test frequency or by
adjusting an optional variable phase shifter or line stretcher. Quadrature
is attained when the meter is set to center scale, zero.
Agilent Technologies E5500 Phase Noise Measurement System 8-11
Residual Measurement Fundamentals
Figure 8-8 on page
8-12.
The Calibration Options