Beginning the
Measurement
CAUTION
Table 13-2
AM Noise using an Agilent/HP 70420A Option 001
1. From the Measurement menu, choose New Measurement
2. When the Perform a New Calibration and Measurement dialog box
appears, click OK.
3. When the Connect Diagram dialog box appears, click on the hardware
down arrow and select your hardware configuration from the pull-down
list.
Confirm your connections as shown in the connect diagram. At this time
connect your UUT and reference sources to the test set. The input
attenuator (Option 001 only) has now been correctly configured based
on your measurement definition.
The Agilent/HP 70420A test set's signal input is subject to the following
limits and characteristics:
Agilent/HP 70420A Test Set Signal Input Limits and Characteristics
Frequency
Maximum Signal Input Power
At Attenuator Output, Operating Level
Range:
• RF Phase Detectors
• Microwave Phase Detectors
Agilent Technologies E5500 Phase Noise Measurement System 13-7
AM Noise Measurement Examples
Limits
50 kHz to 1.6 GHz (Std)
50 kHz to 26.5 GHz (Option 001)
50 kHz to 26.5 GHz (Option 201)
Sum of the reference and signal input
power shall not exceed +23 dBm
0 to +23 dBm (Signal Input)
+15 to +23 dBm (Reference Input)
0 to +5 dBm (Signal Input)
+7 to +10 dBm (Reference Input)
.