Agilent Technologies E5500A User Manual page 250

Phase noise measurement system
Hide thumbs Also See for E5500A:
Table of Contents

Advertisement

Residual Measurement Fundamentals
The Calibration Options
Procedure
1. Connect circuit as per
Figure 8-6
Measuring Power at Phase Detector Signal Input Port
2. Measure the power level that will be applied to the signal input of the
Agilent/HP 70420A's Phase Detector. The following chart shows the
acceptable amplitude ranges for the Agilent/HP 70420A Phase
Detectors.
Table 8-1
Acceptable Amplitude Ranges for the Phase Detectors
50 kHz to 1.6 GHz
Ref Input (L Port)
+
15 dBm
to
+
23 dBm
1. Agilent/HP 70420A Phase Noise Test Set Options 001 and 201
3. Locate the power level you measured on the left side of the Phase
Detector Sensitivity Graph
the graph at the measured level and find the corresponding Phase
Detector constant along the right edge of the graph. This is the value you
will enter as the Current Detector Constant when you define your
measurement. (Note that the approximate measurement noise floor
provided by the Signal Input port level is shown across the bottom of the
graph.)
8-10 Agilent Technologies E5500 Phase Noise Measurement System
Figure
8-6, and tighten all connections.
Phase Detector
Signal Input (R
Ref Input (L Port)
Port)
0 dBm
to
+
23 dBm
(Figure 8-7 on page
1
1.2 to 26.5 GHz
Signal Input (R
Port)
+
7 dBm
0 dBm
to
+
+
10 dBm
8-11). Now move across
to
5 dBm