Beginning the
Measurement
CAUTION
Absolute Measurement Examples
Stable RF Oscillator
Chapter 6, "Absolute Measurement Fundamentals"
determining the effect the amplifiers noise will have on the measured noise
floor.
VCO Reference Source
This setup calls for a second signal source that is a similar type to that of the
UUT. The second source is used as the reference source. In order for the
noise measurement results to accurately represent the noise of the UUT, the
noise level of the reference source should be below the expected noise level
of the UUT. (For additional help in selecting an appropriate reference
source, refer to
Chapter 6, "Selecting a
1. From the Measurement menu, choose New Measurement
2. When the Perform a New Calibration and Measurement dialog box
appears, click OK.
3. When the Connect Diagram dialog box appears, click on the hardware
down arrow and select your hardware configuration from the pull-down
list.
Confirm your connections as shown in the connect diagram. At this time
connect your UUT and reference sources to the test set. The input
attenuator (Option 001 only) has now been correctly configured based
on your measurement definition.
The Agilent/HP 70420A test set's signal input is subject to the following
limits and characteristics:
7-8 Agilent Technologies E5500 Phase Noise Measurement System
for details on
Reference".)
.