Agilent Technologies E5500A User Manual page 277

Phase noise measurement system
Hide thumbs Also See for E5500A:
Table of Contents

Advertisement

Table 9-2
Agilent/HP 70420A Test Set Signal Input Limits and Characteristics
• Microwave Phase Detectors
CAUTION:
To prevent damage to the Agilent/HP 70420A Test Set's hardware components,
the input signal must not be applied to the test set's signal input connector until
the input attenuator (Option 001) has been correctly set by the phase noise
software, which will occur at the connection diagram.
Input Impedance
Figure 9-1
Setup diagram for the Agilent/HP 8349A Amplifier Measurement
Example.
Agilent Technologies E5500 Phase Noise Measurement System 9-9
Residual Measurement Examples
Amplifier Measurement Example
0 to +5 dBm (Signal Input)
+7 to +10 dBm (Reference Input)
Characteristics:
50 ohm Nominal

Advertisement