Table 18-5
Note: Special Function 120 must be enabled for the
DCFM
Model
Band
Numbe
Option
Minimum
r
(MHz)
8664A
2060
8664A
1500
8664A
1030
8664A
750
8664A
515
8664A
375
8664A
257.5
8664A
187.5
8664A
30
8664A
5
8664A
0.05
1
Takes into account limited tuning resolution available in linear FM (Special Function 120, refer to
Functions" on page
18-15).
Agilent/HP 8664A
Mode Keys
Table 18-6
Reference Graphs and Tables
Agilent/HP 8664A Frequency Limits
Agilent/HP 8664A Frequency Limits
Agilent/HP 8664A Frequency Limits
Band
Maximum
Mode 3
(MHz)
3000
400000
2059.99999999
200000
1499.99999999
200000
1029.99999999
100000
749.99999999
100000
514.99999999
50000
374.99999999
50000
257.49999999
25000
187.49999999
200000
29.99999999
100000
4.99999999
Max FM = MIN(Above, Carrier freq - 9 kHz)
The [Mode 2] key provides a median range of FM deviation and RF
output switching time, as shown in the following table.
The [Mode 3] key provides the lowest noise level at the RF output, FM
deviation bandwidth is narrower, and the RF switching time is slower
than in either Modes 1 or 2.
Operating Characteristics for Agilent/HP 8664A Modes 2 and 3
Charateristic
RF Frequency Switching Time
FM Deviation at 1 GHz
Phase Noise (20 kHz offset at 1 GHz)
18-18 Agilent Technologies E5500 Phase Noise Measurement System
1
Minimum Recommended PTR (Peak Tune Range)
PTR =FM Deviation x VTR
Mode 2
10000000
10000000
5000000
5000000
2500000
2500000
1250000
1250000
5000000
5000000
Synthisis Mode
Mode 2
Mode 3
200 ms
350 ms
1 MHz
100 kHz
-130 dBc
-136 dBc