Table A-2
CAUTION
Cleaning Supplies Available from Agilent Technologies
Product
Lint-Free cloths:
Small foam swabs:
Large foam swabs
Do not allow excessive alcohol to run into the connector. Excessive alcohol
entering the connector collects in pockets in the connector's internal parts.
The liquid will cause random changes in the connector's electrical
performance. If excessive alcohol gets into a connector, lay it aside to allow
the alcohol to evaporate. This may take up to three days. If you attach that
connector to another device it can take much longer for trapped alcohol to
evaporate.
Agilent Technologies E5500 Phase Noise Measurement System A-5
Connector Care and Preventive Maintenance
Using, Inspecting, and Cleaning RF Connectors
Part Number
9310-4242
9300-1270
9300-0468