Stable Rf Oscillator; Required Equipment - Agilent Technologies E5500A User Manual

Phase noise measurement system
Hide thumbs Also See for E5500A:
Table of Contents

Advertisement

CAUTION

Required Equipment

NOTE
Absolute Measurement Examples

Stable RF Oscillator

Stable RF Oscillator
This measurement example will help you measure the phase noise of a stable
RF oscillator with frequency drift of <20 ppm over a period of thirty
minutes.
To prevent damage to the Agilent/HP 70420A test set's hardware
components, the input signal must not be applied to the signal input
connector until the input attenuator has been correctly set for the desired
configuration, as show in
when the Connection Diagram appears.
The following equipment is required for this example in addition the phase
noise test system and your unit-under-test (UUT).
To ensure accurate measurements, you should allow the UUT and
measurement equipment to warm up at least one hour before making the
noise measurement.
Table 7-1
Equipment
VCO Reference Source
Coax Cables
7-2 Agilent Technologies E5500 Phase Noise Measurement System
Table 7-3 on page
7-9. Apply the input signal
Required Equipment for the Stable RF
Oscillator Measurement Example
Quantity
Comments
Chapter 6, "Selecting a
1
Refer to
Reference"
about reference source
requirements
And adequate adapters to connect
the UUT and reference source to
the test set.
for more information