AM Noise Measurement Fundamentals
Method 1: User Entry of Phase Detector Constant
Figure 12-10
Modulation Sideband Calibration Setup
4. Measure noise data and interpret the results.
NOTE
The quadrature meter should be at zero volts due to the blocking capacitor at
the AM detector's output.
Agilent Technologies E5500 Phase Noise Measurement System 12-11