Agilent Technologies E5500A User Manual page 265

Phase noise measurement system
Hide thumbs Also See for E5500A:
Table of Contents

Advertisement

Procedure
1. Connect circuit as per
connections. Note that the input signal into the directional coupler is
being supplied to the coupler's output port.
Figure 8-15
Calibration Setup
2. Measure the power level that will be applied to the Signal Input port of
the Agilent/HP 70420A's Phase Detector. The following chart shows the
acceptable amplitude ranges for the Agilent/HP 70420A Phase
Detectors.
Table 8-7
Acceptable Amplitude Ranges for the Phase Detectors
50 kHz to 1.6 GHz
Ref Input (L Port)
+
15 dBm
to
+
23 dBm
1. Agilent/HP 70420A Phase Noise Test Set Options 001 and 201
Agilent Technologies E5500 Phase Noise Measurement System 8-25
Figure 8-15 on page
Phase Detector
Signal Input (R
Ref Input (L Port)
Port)
0 dBm
to
+
23 dBm
Residual Measurement Fundamentals
The Calibration Options
8-25, and tighten all
1
1.2 to 26.5 GHz
Signal Input (R
Port)
+
7 dBm
0 dBm
to
+
+
10 dBm
to
5 dBm