The instrument automatically sets the start wavelength at 1490 nm, stop wavelength at 1590
nm, span, and reference level, and then performs a fully automatic, internal auto align. The input
signal is aligned at equally spaced alignments (minimum 50 nm spacing between points) for the
internal, multi-mode fiber or (minimum 5 nm spacing between points) for the external, single-
mode fiber (Agilent 86146B only).
N o t e
Error 5056, Trajectory align cannot find input signal, will occur if a broadband light source is not
connected to the front-panel input connector.
Error 5060, Trajectory align failed, will occur if the align procedure failed.
Key Path
Related Functions
Remote Commands
Center Wavelength (WL)
Sets the center wavelength using the knob, step keys, or numeric keys. The span remains con-
stant. The center wavelength and the start and stop wavelength settings are related as follows:
Center Wavelength = (stop wavelength + start wavelength)/2
Key Path
Related Functions
Remote Commands
System > More System Functions > Service Menus > Adv Service Functions > More Adv Ser-
vice Menu > Multi-Point Align > Calibrator Multi-Pt Align
Auto Align & Add To Trajectory
Auto Align Preset
User Source Multi-Pt Align
CALibration:ALIGn:INTernal
Wavelength > Center WL
Start WL, Stop WL, Span
SENSe:WAVelength:CENTer
Function Reference
Center Wavelength (WL)
3-17