Agilent Technologies 8614 B Series User Manual page 129

Optical spectrum analyzer
Table of Contents

Advertisement

Power Calibration Setup
Accesses the setup panel to set the calibration power and calibration wavelength.
Set Calibration Power is the exact amplitude of the calibration source that will be used for the
next power calibration. The power entered must be within +10 dB and –3 dB of the amplitude
measured using the factory calibration.
Set Calibration Wavelength is the approximate wavelength of the calibration source that will be
used for the next user power calibration. The wavelength entered must be within 2.5 nm of the
wavelength measured using the factory calibration.
Power Calibration External Path enables power calibration for the external fiber path (path 2).
Calibration for Path 1 is always enabled. A 9 µm patchcord must be connected between the
Monochromator Output and the Photodetector Input before performing a power calibration on
the external path.
Key Path
Remote Commands
Power Display
For 86146B only
In Power Meter mode, turns power display on or off.
Key Path
Remote Commands
System > Calibration > Power Cal Setup
CALibration:POWer
CALibration:WAVelength
Appl's > Measurement Modes > Power Meter Mode > Power Display
DISPlay[:WINDow[1]]:POPup[1|2|3|4][:ALL] OFF|ON|0|1
Function Reference
Power Calibration Setup
3-55