Agilent Technologies 8614 B Series User Manual page 219

Optical spectrum analyzer
Table of Contents

Advertisement

Table 5-7. Application-Specific Warnings (5 of 6)
Error Number
Error Description [description/explanation/examples]
8029
desc = "The requested function is not yet implemented."
help = "The statement within the specification file is not implemented in this
version of the application. The word in parentheses has been reserved for future
use. The statement at the specified line number needs to be removed before the
specification file can be imported."
8030
desc = "A SETUP statement is needed before SWEEP."
help = "The wavelength limits of the SWEEP statement are checked against the
wavelength limits in the SETUP statement. This check requires that the SETUP
statement occur before the SWEEP statement. The line number indicates which
SWEEP statement is not preceded by the SETUP statement."
8031
desc = "The SWEEP limits are outside the SETUP range."
help = "The start and stop values for the SWEEP statement must be within the start
and stop values of the SETUP statement. The line number indicates which SWEEP
statement has values out of range. The number in the parentheses indicates which
parameter is out of range."
8032
desc = "The units of the inputs are not compatible."
help = "The units of the inputs to a math operation need to be compatible with the
operation desired. Addition and subtraction need to have matching units.
Multiplication needs to have a unitless parameter. Division needs a unitless
denominator or the units for the numerator and denominator need to match. The
line number indicates which statement had the mismatched input units."
8033
desc = "The reference point is outside the SETUP range."
help = "The reference point value must be within the start and stop values of the
SETUP statement. The line number indicates which statement had the value out of
range."
8034
desc = "The values are associated with different axes."
help = "The variables used in a basic math operation like ADD or DIV must be
associated with the same data axis (e.g. wavelength or amplitude). Mixing the axes
values within a single math operation is not allowed. The line number indicates
which statement had the incompatible variables."
8035
desc = "The search limits are outside the SETUP range."
help = "The search limits for statements like PEAK or CENTER_OF_MASS must be
within the start and stop values of the SETUP statement. The line number indicates
which statement has values out of range. The number in the parentheses indicates
which parameter is out of range."