Recall Menu
Accesses the Recall Setup panel for the following settings:
• Recall (all measurement data or trace data only)
• Recall From (recall from a floppy disk or from the internal drive
• Network File Share 1|2|3|4
• Network File Path
N o t e
To insure accurate measurements, a wavelength calibration should be performed each time
measurement data is recalled from memory.
Key Path
Remote Commands
Reference Level
Specifies the reference level value, the maximum expected power to be measured, at the refer-
ence level position. The reference level position is indicated on the display by a dashed green line
and the REF annotation on the display. The default position is one major graticule division from
the top of the display. The reference level is entered in dBm. Use the Amplitude Setup panel to
change the reference level position.
page 3-64
Key Path
Related Functions
Remote Commands
Save/Recall > Recall Menu
MMEMory:LOAD:TRACe
Amplitude > Reference Level
Amplitude Setup (Reference Level Position), Scale/Div
DISPlay:WINDow:TRACe:Y:SCALe:RLEVel
See "Reference Level Position (Amplitude Setup Panel)" on
Function Reference
Recall Menu
3-63