Using the Instrument
Analyzing Measurement Data
Using Trace Math to Measure Wavelength Drift
1 From the front panel, press Traces > Active Trace > TrA.
2 Press Single Sweep, Bandwidth Sweep, Single Sweep to update Trace A then press Traces,
Update A off.
3 Press Active Trace > TrB.
4 Press Sweep > Repeat Sweep On to continuously update the measured response on Trace B.
5 Press Traces > Trace Math, Default Math Trace C > Log Math C = A – B.
You can now monitor the wavelength drift of your device over time.
Also
Refer to "Log Math C=A–B" on page
Math F=C–D" on page
3-32.