Agilent Technologies 8614 B Series User Manual page 21

Optical spectrum analyzer
Table of Contents

Advertisement

To perform an Auto Align
For maximum amplitude accuracy, perform an automatic alignment whenever the optical spec-
trum analyzer has been moved, subjected to large temperature changes, or following warm-up.
See
"Auto Align" on page 3-9
1 Connect a fiber from the source to the input connector of the optical spectrum analyzer. Be sure
to follow the good connector practices described in
Measurements" on page
2 Enable the source. Press Markers > Peak Search to find the peak signal power.
3 Press the front-panel Auto Align key to optimize the detection of the incoming signal. This takes a
few moments to complete.
To perform a peak search
4 Press the front-panel Auto Meas key to locate and zoom-in on the signal. Please wait until the Auto
Measure routine is complete. A marker is placed on the peak of the displayed signal.
Trace with normal marker.
To zoom in on the signal
Press the Span softkey and then use the knob, step keys, or numeric keypad to zoom in on the
signal.
for more information.
"Cleaning Connections for Accurate
6-8.
Getting Started
Making a Measurement
1-13