Agilent Technologies 8614 B Series User Manual page 157

Optical spectrum analyzer
Table of Contents

Advertisement

N o t e
The auto align now will overwrite any previous align data. To preserve current auto align data,
select Switch Path No Auto Align.
The data returned by the alignment is stored in both the 9 µm and the 50 µm table. With the data
stored on both tables, the centering for the 50 µm path is improved due to the increased resolu-
tion bandwidth of the 9 µm path. Once the align is complete or if you select No Auto Align, the
instrument will be ready to detect data via the 50 µm path.
Key Path
Related Functions
Remote Commands
Switch Path No Auto Align
For 86146B Filter Mode only.
Selects the 9 µm filter mode path. You should select this function if the instrument has not be
moved, subjected to temperature changes >2° C, turned off and warmed up for at least an hour
at the start of each day or to preserve previous align data.
Key Path
Related Functions
Remote Commands
Appl's > Measurement Modes > Filter Mode > Switch Path Auto Align Now
Switch Path No Auto Align
ROUTe:PATH:EXTernal
CALibration:ALIGn:FILTer
Appl's > Measurement Modes > Filter Mode > Switch Path Auto Align Now
Switch Path Auto Align Now
ROUTe:PATH:EXTernal
Function Reference
Switch Path No Auto Align
3-83