Table 5-6. OSA Warnings (11 of 12)
Error Number
Error Description [description/explanation/examples]
6749
desc = "Peak not found"
help = "A peak search was requested. There are no valid peaks on the current
marker's trace. Try decreasing the marker peak excursion setting or adjust the
sensitivity setting."
6750
desc = "Pit not found"
help = "A pit search was requested. There are no valid pits on the current marker's
trace. Try decreasing the marker pit excursion setting or adjust the sensitivity
setting."
6751
desc = "Trace has no centroid"
help = "The trace has no centroid. This may occur because the sum of trace points'
amplitudes is zero or there a no trace points."
6752
desc = "The reference point is outside trace bounds"
help = "The reference point's wavelength is either too small or too large. The
reference point will be clipped to a trace endpoint."
6753
desc = "The window specified is invalid."
help = "The window specified is invalid. The window should be a constant."
6754
desc = "Search failed: no data"
help = "A marker search operation was requested. After clipping the trace data to
screen limits and line marker limits (if enabled) there was no data to search. This
can be caused when all trace data points are beyond current X axis screen limits or
when there are no trace data points between the line markers. Trace data values
which exceed the current Y axis screen limits will not cause this problem. Adjust the
current X axis screen limits and/or move the line markers to include at least one
trace data point."
6755
desc = "Attempt to divide by 0"
help = "At least one point was attempted to be divided by 0. The result for these
divisions has been set to not-a-number. Please realize that further calculations with
not-a-number values are undefined."
6756
desc = "Syntax error; bad token: "
help = "The grammar expression entered cannot be parsed. Please check the string
entered. The bad token attempts to indicate where the error occurred. The end of
line indicates that the OSA expected more information. Please consult the manual
for additional help."
7998
desc = "Unknown error detected"
help = "An unlisted error was reported by the instrument software. If this error
persists contact Agilent Technologies for assistance."
Status Listings
OSA Warnings
5-27