Maintenance
Cleaning Connections for Accurate Measurements
To clean the recessed end of the connector saver
• To clean the recessed end of the connector saver, it is recommended that a wrapped tip swab or
stick cleaner be used. Berkshire's LT670183 wrapped tip cotton swab, or Cletop stick cleaners
have proven to be an effective cleaning solution. To clean the recessed end of the connector saver,
insert the stick type cleaner, and rotate it server times the same direction. Each swab should only
be used once. While this may not remove all of the contaminants, it will displace them away from
the center of the fiber (away from the core).