Table 5-6. OSA Warnings (5 of 12)
Error Number
Error Description [description/explanation/examples]
5051
desc = "AutoMeasure cannot find input signal at marker"
help = "The auto-measure function was requested. The auto-measure at marker
option was enabled but the active marker was not placed on a valid signal. A valid
signal was found but the active marker is too far from that signal. To correct this
problem place the marker on the desired signal turn off all markers or disable the
auto-measure at marker function."
5052
desc = "Sweep time auto is not allowed in zero span"
help = "The sweep time function was requested to switch from manual to auto
while in zero span. Automatic sweep time coupling cannot be done in zero span.
The desired sweep time must be set explicitly."
5053
desc = "Noise marker not allowed."
help = "A request was made to enable the noise marker readout. The readout
cannot be enabled for one of the following reasons. Noise markers cannot be used
on traces which have X axis units other than meters. This typically occurs in zero
span where the X axis units are seconds. Noise markers cannot be used on traces
which have Y axis units other than power. This typically happens when trace math
is on and the trace Y axis represents a ratio."
5054
desc = "Bandwidth markers are not allowed in zero span"
help = "A request was made to enable the bandwidth marker readout while the
OSA was in zero span. Bandwidth markers are not permitted when span is set to
zero. If a bandwidth marker readout is required set span to a value greater than
zero."
5055
desc = "Firmware Upgrade was not successful"
help => "A firmware upgrade operation was requested. The firmware upgrade
operation cannot be done at this time due to an internal software problem. " + <-
.serviceCenterHelp
5056
desc = "Trajectory align cannot find input signal"
help = "The trajectory align procedure cannot find a usable input signal. Make sure
you have a signal connected to the optical input."
5057
desc = "Invalid settings for trajectory align"
help = "The external trajectory align function was requested. It cannot be executed
because the start and/or stop wavelength settings are invalid. Start and stop
wavelengths must be between 600nm and 1700nm. In addition the difference
between start and stop wavelengths (span) must be greater than 25nm. To correct
this problem adjust the start and/or stop wavelength settings to be valid."
Status Listings
OSA Warnings
5-21