Function Reference
Calibration
Key Path
Related Functions
Remote Commands
Calibration
Accesses the Power Cal Setup and Wavelength Calibration Setup softkeys. The Power Calibra-
tion Setup panel will show the date of the last factory performed power calibration and the date
of the last successful user-performed power calibration. It also allows you to set the calibration
power and wavelength used for the next calibration.
ments" on page 2-13
Set Calibration Power is the exact amplitude of the calibration source that will be used for the
next power calibration. The power entered must be within +10 dBm and –3 dBm of the ampli-
tude measured using the factory calibration.
Set Calibration Wavelength is the approximate wavelength of the calibration source that will be
used for the next user power calibration. The wavelength entered must be within 2.5 nm of the
wavelength measured using the factory calibration.
Pressing the Perform Calibration softkey will display on-screen instructions for performing the cal-
ibration.
Key Path
Remote Commands
Calibrator Multi-Pt Align
Adjusts the mechanical position of the instrument's internal optical components ensuring ampli-
tude accuracy of your measurements. Before initiating the alignment, connect the internal cali-
brator to the front-panel input connector.
on page 2-16
3-16
Markers > Marker Setup > BW Marker Units
Marker BW
CALCulate[1|2|3|4|5|6]:MARKer[1|2|3|4]:FUNCtion:BANDwidth:READout
Systems > Calibration
CALibration:POWer:POWer
CALibration:POWer:Wavelength
Refer to "Calibrating Wavelength Measure-
Refer to "External Multipoint Wavelength Calibration"