Function Reference
Reference Level Position (Amplitude Setup Panel)
Reference Level Position (Amplitude Setup Panel)
Determines the position of the reference level on the graticule. Setting this value to zero divisions
places the reference level on the very bottom of the grid. Setting the reference level to 10 divi-
sions places the reference level at the top of the grid. The default (or preset) position is 9 divi-
sions, which is one major graticule down from the top of the
page 3-63
The reference level position can be changed using the knob, step keys, or numeric keypad.When
using the knob or step keys, the value will automatically change. When using the numeric key-
pad, a terminator must be selected from the softkey panel.
Key Path
Related Functions
Remote Commands
Remote File Share Panel
Uses the LAN (local area network) to store, recall, or delete data to remote hard drives. The data
can then be accessed and shared among the users and printed on designated, network printers.
See "Recalling Measurement and Trace Data" on page 2-23
N o t e
To access the file and printer share softkeys, you must first configure the network and enter the
user share identity/user profile information for remote shares. The softkeys for file and printer
share will then become available for selection.
Key Path
Related Functions
Remote Commands
3-64
Amplitude > Amplitude Setup > Reference Level Position
Amplitude Units, Amp Display Mode, Reference Level, Scale/Div
DISPlay:WINDow:TRACe:Y:SCALe:RLPOS
System > More System Functions > GPIB & Network Setup > File Shares > Remote File
Shares
Remote Printer Share
MMEMORY:FSHAre [1|2|3|4][:PATH] <param>
MMEMORY:FSHAre [1|2|3|4][:ADDRess]<param>
display.See "Reference Level" on