Resistance Setting - R&S NGM202 User Manual

Hide thumbs Also See for NGM202:
Table of Contents

Advertisement

®
R&S
NGL200/NGM200
Query parameters:
<current>
Example:

7.5.5 Resistance Setting

The SOURce:RESistance subsystem contains the commands for setting the resist-
ance limit of the output channels. The default unit is ohms.
User Manual 1178.8736.02 ─ 06
MAX
Depending on the set voltage level, the maximum set current is
6.010 A.
For voltage range from 0 V to 6 V, maximum set current is 6.01
A.
For voltage > 6 V, maximum set current is 3.01 A.
UP
Increases current by a defined step size. See
]CURRent[:LEVel][:IMMediate]:STEP[:INCRement]
on page 106.
DOWN
Decreases current by a defined step size. See
]CURRent[:LEVel][:IMMediate]:STEP[:INCRement]
on page 106.
MIN | MAX
MIN
Returns minimum current.
MAX
Returns maximum current.
See
Example "Configuring the current output"
Remote Control Commands
Configuration Commands
[SOURce:
[SOURce:
on page 105.
107