6 Instrument Functions

6.1 Setting the Channels Voltage and Current

The R&S NGL/NGM comes with the following instrument models:
Models
NGL201, NGM201
NGL202, NGM202
Toggle the respective channel key ([Ch1], [Ch2] ) on the front panel to select these
channels. When a channel is selected, the respective channel key illuminates.
Figure 6-1: Ch2 key illuminates when selected
Set output voltage and current
Voltage, current settings
If "Arbitrary" function of a selected channel is enabled, the respective channel voltage
or current setting is disabled.
See
The R&S NGL/NGM adjusts the voltage values between 0 V to 20.05 V with a step
size of 1 mV and current values between 0 A to 6.01 A with a step size of 0.1 mA.
The setting of current value corresponds to the I
advisable to set the current limit before operating the instrument to prevent damage to
the load and instrument in the case of malfunction such as a short-circuit.
1. Press [Home] key.
The R&S NGL/NGM displays the home window.
2. Set voltage or current in the home window.
The R&S NGL/NGM displays the on-screen keypad to set value.
3. Enter the required voltage or current value.
4. Confirm value with the unit key (V/mV or A/mA).
5. Press the required channel key ([Ch1] or [Ch2]) on the front panel.
The selected channel key is illuminated. See
6. Press the [Output] key on the front panel.
User Manual 1178.8736.02 ─ 06
"Channel settings"
on page 62.
Instrument Functions
Setting the Channels Voltage and Current
Channels
Ch1
Ch1, Ch2
of the respective channel. It is
max
Figure
6-1.