Channel Display Area - R&S NGM202 User Manual

Hide thumbs Also See for NGM202:
Table of Contents

Advertisement

®
R&S
NGL200/NGM200
Function
"Output Delay"
Sense connection

5.1.2 Channel Display Area

The R&S NGL/NGM displays two channels display area (Ch1, Ch2) for NGL202,
NGM202 and a single channel display area (Ch1) for NGL201, NGM201. The respec-
tive channel settings and functions are displayed for each channel.
Figure 5-2: Channel display area for 2-channel model
1 = Output power displays in watt
2 = Source output resistance/emulated internal impedance displays in ohms
3 = "Settings" button opens instrument device/channel menu window
4 = "Expand/Collapse" button toggles between home window and channel overview window
5 = Output voltage displays in volt. The display resolution for voltage is five digits after the decimal point
6 = Set voltage level. This level is limited by the set value defined in "Safety Limits"
7 = Output current displays in ampere. The display resolution for current is five digits after the decimal point
8 = Set current level. This level is limited by the set value defined in "Safety Limits"
9 = DVM function (available only with R&S NGM power supply series)
Operating mode
Different font colors on the screen are used to differentiate the various output status
and operating condition of the instrument. It is easy to know and confirm the different
output status and operating conditions of the instrument by looking at the colors.
User Manual 1178.8736.02 ─ 06
Description
If enabled, the icon is highlighted in white.
The delay is the time between activation of the out-
put and applying voltage to the output.
See
Chapter 6.2.3,
"Output",
If sense connection is detected, the icon is highligh-
ted in white.
Operating Basics
Display Overview
on page 44.
29