2. Set the required value.
The R&S NGL/NGM displays the onscreen keypad for entry.
3. Confirm value with the unit keys.
The output impedance control loop depends on the setting "Fast Transient Response".
If the setting is "ON", the focus is on control loop speed, which is recommended for
resistive loads. Otherwise, the focus is on stability and should be preferred for capaci-
tive or inductive loads. See
The output impedance is only active during CV (constant voltage) operation mode.
6.2.3.2

Delay

1. Select the "Delay" menu item to configure the required values.
The R&S NGL/NGM displays the "Output Delay" dialog.
2. Set the required value.
The R&S NGL/NGM displays the onscreen keypad for entry.
3. Confirm value with the unit keys.
The output delay is the time between the "Output On" event and the available voltage
at the output terminals. See
User Manual 1178.8736.02 ─ 06
Chapter 6.2.2, "Fast Transient
Figure
6-7.
Instrument Functions
Activating the Channels Output
Response",
on page 43.